PBL Logo - click here to return to homepage spacer
Get QuoteSubmit SamplesTest and Report spcr
 
 
Home >> Medical Device Services >> Validation/Manufacturing Support >> Sterilization Validations >> Methods and Requirements
STERILIZATION VALIDATIONS: Methods and Requirements

Radiation Sterilization | Ethylene Oxide Sterilization | Steam Sterilization | Chemical Sterilization
 

STEAM STERILIZATION

Manufacturers of Reusable Medical Devices designed to be sterilized by steam must provide detail instructions for reprocessing. The manufacturer shall validate that the instructions provided are capable of sterilizing the medical device for its intended use. The validated packaging configuration as well as steam sterilization parameters must be provided to the health care facilities in these instructions.

The most common steam sterilization cycles used in the industry are the gravity-displacement and dynamic-air-removal. In gravity-displacement the steam enters the sterilizer chamber and displaces the residual air through an open vent. A more efficient cycle type is the dynamic-air-removal, specifically the prevacuum cycle, in which the air is removed from the chamber and load by means of pressure and vacuum excursions. Flash sterilization cycles consist of higher temperatures for a shorter time and are designed for processing patient care items for immediate use. At Pacific BioLabs the overkill sterilization model is used with these two cycle types to demonstrate a 10-6 Sterlity Assurance Level (SAL). For more information see ANSI/AAMI ST79:2006 Comprehensive guide to steam sterilization and sterility assurance in health care facilities.

The following are the most common steam sterilization cycles validated at PBL:

Cycle Type Packaging Temp (°C/°F) Minimum Exposure Time (minutes)
Gravity-Displacement           Wrapped Instruments 121/250 30 minutes
Gravity-Displacement           Wrapped Instruments 132/270 15 minutes
Gravity-Displacement           Wrapped Instruments 135/275 10 minutes
Gravity-Displacement Flash Unwrapped Non-Porous Instruments 132/270  3 minutes
Gravity-Displacement Flash Unwrapped Non-Porous Instruments 135/275 3 minutes
Prevacuum Flash Wrapped Instruments 132/270 4 minutes
Prevacuum Flash Wrapped Instruments 135/275 3 minutes
Prevacuum Flash Unwrapped Non-Porous Instruments 132/270 3 minutes
Prevacuum Flash Unwrapped Non-Porous Instruments 135/275 3 minutes